2024年11月13日
The post-market surveillance statutory instrument was laid in Parliament on 21 October 2024 to amend the Medical Devices Regulations 2002 (the UK MDR). This draft legislation seeks to insert new post-market surveillance (PMS) requirements for medical devices placed on the market in Great Britain (GB). If passed by Parliament, it is expected that this legislation will come into force in summer 2025, following a six-month implementation period. This article has been updated following publication of the MHRA of their response to WTO comments on 3 July 2024, and again following the statutory instrument being laid in Parliament on 5 November 2024.
The publishing of this draft statutory instrument follows the consultation on the future regulation of medical devices in the UK and is part of the wider revamp of the UK's medical device regulatory framework following Brexit. Medical devices placed on the market in Northern Ireland will continue to be subject to the EU regulatory framework under the Northern Ireland Protocol.
The requirements in the proposed PMS regulations closely resemble those already established in the EU Medical Device Regulation 2017/745 (the EU MDR) and the EU In Vitro Diagnostic Medical Devices Regulation 2017/746 (the EU IVDR). However, there are differences, some of which are subtle, and some significant, but which nevertheless mean that the requirements will not be the same in GB as in the EU. Therefore, if the draft legislation becomes law in this form, manufacturers across Europe should check their planned PMS activities specifically against the UK legislation and not automatically assume that they are the same as in the EU.
Naturally, reports for devices placed on the market or put into service in GB must be submitted to the MHRA and in some cases to the UK approved body and UK responsible person (UKRP), if any. If the MHRA is separately notified of incidents and informs the manufacturer of the complaints, the manufacturer must determine if it is a "serious incident" and if so, must take action accordingly and if not, explain why. Similarly, if the MHRA notifies a manufacturer of an identified risk or safety concern, then the manufacturer is obliged to investigate and report back to the MHRA on its conclusion and provide any corrective and preventative action (CAPA). The MHRA has clarified in its response to WTO comments on the draft statutory instrument that if the MHRA determines that an incident is reportable, there is no process for appealing that decision. Judicial review would therefore be the only means to do so.
There are some PMS aspects which are the same as the EU. These include the requirement to maintain a PMS system based on the PMS plan for each device placed on the market. As in the EU, this system is to be used to identify preventive and corrective actions. However, the UK has chosen to include many subtle differences which will require careful analysis when compared with an EU PMS strategy to ensure that the UK regulations are respected for devices placed on the market or put into service in GB.
In its response to the WTO, the MHRA explained that it has decided on divergent definitions from the EU for the purposes of clarity or global alignment, looking to the IMDRF for inspiration. The MHRA has also defended its prescriptive requirements for data collection as being necessary to ensure a consistent level of stricter safety standards.
The GB definition of "serious incident" includes the same terminology as the EU regulations, but then extends it to include a definition of "serious deterioration of any person's state of health" which are defined as: : life-threatening illness or injury, permanent impairment of a body structure or body function, hospitalisation or prolongation of hospitalisation, medical treatment, chronic disease, and foetal distress or foetal death. This potentially expands the types of serious incidents captured by the definition.
Reporting and investigation of serious incidents: the reporting requirements match the EU ones, except on a few points:
The exception of expected and documented side-effects does not appear in the UK legislation. These must therefore be reported in GB where they might not be in the EU.
The WTO considered that under-reporting of incidents to manufacturers is not addressed. This would require legal obligations on those other than manufacturers, such as HCPs and HCOs to report, which has never been considered as an obligation to be included in UK law. We note however the recent pro-active approach by the MHRA in directing patients to report safety problems with continuous glucose monitors or insulin pumps.
Preventive and corrective actions: manufacturers should of course note the potentially broader definition of "preventive action", and that the notification is to the UKRP or approved body (if either exist) and the MHRA. The provisions are otherwise in parallel with the EU regulations. The WTO noted the additional reporting requirement as being divergence from EU, but the MHRA consider this as a means to update their Approved Body (if of course they have one).
Post-market surveillance report (PMSR): the requirements are the same as EU regulations, save that:
Each PSUR must be submitted to the approved body (if there is one) for review if the device is: (i) class III; (ii) implantable (including active implantable); or (iii) listed in Annex II of the in vitro diagnostic directive 98/79 (IVDD). The approved body must then issue a report to the manufacturer and any UKRP setting out its conclusions.
"required risk analysis" has been clarified by the MHRA as being a systematic analysis of hazards and an estimation of risk that is undertaken before conducting a benefit-risk determination.
The MHRA has also clarified that it will require where appropriate for a device that "patient and public engagement" (PPE) is engaged in to identify improvements to meaningful outcomes for patient safety and public health.
Similar incidents occurring outside Great Britain must be reported at the final reporting stage.
Field safety corrective actions and field safety notices: the requirements are mostly the same as for the EU, save that in the original version of the legislation, the manufacturer was to publish the field safety notice on the manufacturer's website and take all reasonable steps to ensure the field safety notice is sent to all known device users. The GB regulations require the manufacturer to take all reasonable steps to ensure the field safety notice is sent to users of the device. The field safety notice must also set out the details of the UKRP.
Trend reporting: these requirements are similar to the EU regulations, albeit with slightly different language, and a detailed list of information to submit. The manufacturer must make these reports to the MHRA via the MORE portal. It is advisable for manufacturers to check the GB list of information before submitting a trend report. The manufacturer is required to respond to MHRA requests with respect to any investigation of the trends within three working days. Despite this being considered unreasonable in the WTO response, the MHRA has kept to this timeline, but adding guidance to allow for flexibility in assessing the appropriate timeframe. The regulation also requires manufacturers to inform the MHRA of FSCA's outside the GB market for the same model of device.
Retention of PMS documentation: the manufacturer and the UKRP (if there is one) must retain documentation until the later of: (i) the PMS period for that device model (see above); or (ii) 15 years in the case of an implantable device, or ten years in the case of any other device. This is different from the EU regulations for which the requirement is "10/15 years after the last device has been placed on the market". This makes the 10/15 years an absolute number and the period of use the more definitive figure for devices with greater longevity. For devices with an expected "lifetime" beyond the 10/15 years (as applicable), the retention period will therefore be longer in GB. Manufacturers of these devices will have to consider what this period is and justify it in their PMS systems for GB.
Once this legislation becomes law, manufacturers must prepare to take on stricter PMS obligations than those currently in place from summer 2025 for medical devices on the market in GB. This will involve amending any current PMS systems, developing PMS plans, and having procedures in place to promptly report and act on incidents, corrective actions and trends in accordance with the GB requirements. Any PMS policies procedures already developed to handle the EU MDR or IVDR will need to be reviewed in light of the GB law. Amendments are likely to be necessary to meet the differing GB requirements to ensure future compliance for devices placed on the market or put into service in GB.